The following tables list our current QA equipment. "Since" means the date of the installation of the actual device or of the most recent software version. Only equipment that we actually use is in the list (former equipment can be found here). For calibrated chambers, we also list the current 60Co calibration factor, which is valid for two years for a certain electrometer/chamber combination. |
Dosemeter | Type | S/N | Since | Last Verified | Remarks |
---|---|---|---|---|---|
UNIDOSwebline | T10021 | 00011 | 2005 | 19.06.2024 | First webline in Austria! |
UNIDOS | T10001 | 10396 | 1996 | 19.06.2024 | |
UNIDOS | T10001 | 10398 | 1996 | 19.06.2023 |
Detector | Type | S/N | Since | Last Verified | Last Calibrated | Dosemeter | Current Nw (60Co) |
---|---|---|---|---|---|---|---|
Roos | TM34001 | 2456 | 2014 | 19.06.2024 | 19.06.2024 | 10396 | 1.002 ± 0.014 |
19.06.2024 | 19.06.2024 | 00011 | 1.001 ± 0.014 | ||||
Roos | TM34001 | 2457 | 2014 | 19.06.2023 | 19.06.2023 | 10398 | 1.003 ± 0.014 |
0.3 ccm | TM31003 | 0939 | 1996 | 19.06.2024 | 19.06.2024 | 10396 | 1.033 ± 0.014 |
19.06.2024 | 19.06.2024 | 00011 | 1.033 ± 0.014 | ||||
0.3 ccm | TM31013 | 0053 | 2003 | 19.06.2023 | 19.06.2023 | 10398 | 1.019 ± 0.014 |
Farmer | TM30013 | 1572 | 2005 | 19.06.2024 | 19.06.2024 | 10396 | 1.013 ± 0.014 |
19.06.2024 | 19.06.2024 | 00011 | 1.014 ± 0.014 | ||||
Farmer | TM30013 | 1582 | 2005 | 19.06.2023 | 19.06.2023 | 10398 | 1.015 ± 0.014 |
Farmer | TM30010 | 2591 | 2014 | 19.06.2024 | 19.06.2024 | 10396 | 1.009 ± 0.014 |
19.06.2024 | 19.06.2024 | 00011 | 1.008 ± 0.014 | ||||
Farmer | TM30010 | 2592 | 2014 | 19.06.2023 | 19.06.2023 | 10398 | 1.012 ± 0.014 |
Device/Software | Type/Version | S/N | Since | Remarks |
---|---|---|---|---|
STARCHECKmaxi | T10033 | 000111 | 2013 | Ionization chamber array for real-time beam measurements |
Detector Interface 4000 | T16039 | 000820 | 2013 | Interface for STARCHECKmaxi |
FIELDCHECK | T40052 | 00011 | 2010 | Checks light/radiation field (used with STARCHECKmaxi) |
FIELDCHECK adapter for MAXI | T10033.1.150 | 00003 | 2011 | For centering FIELDCHECK on STARCHECKmaxi |
BQ-CHECK test object | T42030 | 00165 | 2010 | Beam quality test object (used with OCTAVIUS Detector 1500) |
OCTAVIUS Detector 1500 | T10044 | 111346 | 2015 | 2D Ionization Chamber Array (1405 ionization chambers) |
Detector Interface 4000 | T16039 | 000120 | 2011 | Interface for OCTAVIUS Detector 1500 or 729 |
OCTAVIUS 4D control unit | T40058 | 000875 | 2018 | Part of OCTAVIUS 4D system (large field capability) |
OCTAVIUS 4D modular rotation unit | T40063 | 141563 | 2015 | Part of OCTAVIUS 4D system |
OCTAVIUS top Standard | T40063.1.004 | 141589 | 2015 | OCTAVIUS top with 320 mm diameter |
OCTAVIUS top SRS | T40063.1.002 | 141558 | 2015 | OCTAVIUS top with 170 mm diameter |
OCTAVIUS top LINAC QA | T40063.1.003 | 141562 | 2015 | OCTAVIUS top for linac QA (50 mm buildup) |
OCTAVIUS 4D Inclinometer | T43036 | 132151 | 2018 | Wireless Bluetooth inclinometer for OCTAVIUS 4D |
OCTAVIUS Detector 729 | T10040 | 000118 | 2011 | 2D Ionization Chamber Array (729 ionization chambers) |
Positioning phantom | T40049 | 00083 | 2010 | PMMA phantom to facilitate 5 mm shifts of OCTAVIUS Detector 729 |
OCTAVIUS phantom | T40051 | 00072 | 2010 | For "static" plan verification with the OCTAVIUS Detector 729 |
VeriSoft | V 8.1 | 2022 | Software for 2D-ARRAY evaluation | |
MultiCheck | V 3.7.2 | 2022 | Software for linac QA | |
Epson Expression Scanner | 10000XL | 27168 | 2010 | Scanner (A3 format) for Gafchromic films |
QUICKCHECKwebline | T42031 | 001933 | 2016 | Morning Checkout device, TrueBeam BLUE |
Compensator for 6FFF | T42031.1.310 | 120531 | 2012 | For use with unflattened 6FFF beams, TrueBeam BLUE |
Compensator for 10FFF | T42031.1.320 | 120532 | 2012 | For use with unflattened 10FFF beams, TrueBeam BLUE |
QUICKCHECKwebline | T42031 | 001918 | 2016 | Morning Checkout device, TrueBeam GREEN |
Compensator for 6FFF | T42031.1.310 | 120536 | 2012 | For use with unflattened 6FFF beams, TrueBeam GREEN |
Compensator for 10FFF | T42031.1.320 | 120536 | 2012 | For use with unflattened 10FFF beams, TrueBeam GREEN |
QuickCheck | T42007 | 572 | 2006 | Old Morning Checkout device |
QuickCheck | T42007 | 573 | 2006 | Old Morning Checkout device |
QCWin | V 2.21 | 2006 | Software for QuickCheck |
Device/Software | Type/Version | S/N | Since | Remarks |
---|---|---|---|---|
Linear Array LA48 | 34009 | 00294 | 2002 | |
Interface ME48 | 10006 | 00092 | 1996 | used with Multidos S/N 150 and MEPHYSTO mcc software |
ME48F Extender | 10012 | 00002 | 2001 | used for LA48 in fast measurement mode (100 profiles/second) |
ME48F Interface | 16011 | 00005 | 2001 | used for LA48 in fast measurement mode |
me48f-Soft | V 1.03 - T | 1999 | Software used for LA48 in fast measurement mode | |
AirScanner | T41008.1.040 | 00022 | 2000 | Mounting device for LA48 |
Dosemeter | Type | S/N | Since | Remarks |
---|---|---|---|---|
Multidos | 10004 | 00150 | 1996 | for QuickCheck, in-Vivo dosimetry and Linear Array |
Multidos | 10004 | 00466 | 2001 | for QuickCheck and in-Vivo dosimetry |
Multidos | 10004 | 00467 | 2001 | for QuickCheck and in-Vivo dosimetry |
Detector | Type | S/N | Since | Remarks |
---|---|---|---|---|
Semiflex 3D | TM31021 | 144965 | 2024 | Standard accessory of BEAMSCAN phantom |
Semiflex 3D | TM31021 | 144966 | 2024 | Standard accessory of BEAMSCAN phantom |
0.125ccm | TM31010 | 004882 | 2012 | |
0.125ccm | TM31010 | 004883 | 2012 | |
0.125ccm | TM31002 | 0367 | 1996 | |
Photon Diode | TM60008 | 0011 | 2001 | shielded p-type Si diode |
Photon Diode | TM60008 | 0012 | 2001 | shielded p-type Si diode |
Electron Diode | TM60017 | 000663 | 2012 | unshielded p-type Si diode |
Electron Diode | TM60017 | 000664 | 2012 | unshielded p-type Si diode |
Electron Diode | TM60012 | 0206 | 2006 | unshielded p-type Si diode |
Electron Diode | TM60012 | 0207 | 2006 | unshielded p-type Si diode |
PinPoint 0.015 cm3 | TM31014 | 001079 | 2013 | |
PinPoint 0.015 cm3 | TM31014 | 001080 | 2013 | |
PinPoint 0.030 cm3 | TM31015 | 000491 | 2013 | |
PinPoint 0.030 cm3 | TM31015 | 000492 | 2013 | |
PinPoint3D | TM31016 | 00620 | 2011 | |
PinPoint3D | TM31016 | 0084 | 2006 | |
microLion | TM31018-1,3 | 0207 | 2010 | |
HV-Supply | T16036 | 0063 | 2010 | up to 950V, compatible with UNIDOSwebline and Tandem T10011 |
Diamond Detector | TM60003 | 1034 | 2011 | natural diamond detector |
microDiamond | TM60019 | 122152 | 2013 | synthetic single crystal diamond detector (SCDD) |
microDiamond | TM60019 | 122153 | 2013 | synthetic single crystal diamond detector (SCDD) |
Device/Software | Type/Version | S/N | Since | Remarks |
---|---|---|---|---|
BEAMSCAN Control Unit | TM41047 | 152060 | 2024 | Part of BEAMSCAN water phantom |
BEAMSCAN PMMA Tank | TM41046 | 160713 | 2024 | Part of BEAMSCAN water phantom |
BEAMSCAN Lift Carriage with Built-in Water Reservoir | TM41043 | 151690 | 2024 | Part of BEAMSCAN water phantom |
Water phantom MP3 | 4316 | 00242 | 1996 | The big one |
Tandem | T10011 | 10117 | Electrometer for MP3 | |
Controller | 41013 | 00198 | Controller for MP3 | |
Mephysto mcc | V 3.4 | 2012 | Software for MP3 | |
Small water phantom | 41023 | 00040 | 2005 | horizontal beam, manual operation |
Device/Software | Type | S/N | Since | Remarks |
---|---|---|---|---|
ESTRO Mini Phantom | T40036.1.020 | 2010 | For 0.3 ccm chamber, radial irradiation | |
CIRS Cube Phantom | 002009 | B4007 | 2010 | Cube 20x20x20 cm, holds Farmer chamber and film |
CIRS Radiosurgery Head Phantom | 605 | L1496 | 2010 | Anthropomorphic head with GafChromic film stack |
CIRS Electron Density Phantom | 062 | 2006 | QA of electron density calibration curves | |
QUASAR Penta-Guide | 100-1009 | 0763 | 2010 | QA of IGRT (3D-3D matching and more) |
PTW Inhomogeneity Phantom | T40037 | 2013 | For OCTAVIUS Detector729 (inserts: Lung, Tissue, PMMA, Bone) | |
Civco Iso-Align Phantom | CIV-MTIAD1 | 2014 | used for kV-/MV-scale checks according to ONORM S5290-3 | |
PTW RUBY Head Set Multimet | L981654 | 2021 | Head phantom for simultaneous measurement with 3 detectors | |
T40072.1.800 | 200819 | Main component of set: RUBY head phantom | ||
T40072.1.500 | 200402 | Main component of set: MultiMet Insert | ||
PTW RUBY Set All-inclusive | L981635 | 2020 | Phantom for Machine QA and Patient QA | |
T40072 | 190912 | Main component of set: RUBY Phantom |
Device/Software | Type/Version | S/N | Since | Remarks |
---|---|---|---|---|
QUASAR Respiratory Motion Platform | 100-1010 | 151 | 2010 | Moves platform according to programmed waveform |
QUASAR CATPHAN Shaker | 100-1014 | 626 | 2010 | Moves CATPHAN according to programmed waveform |
QUASAR Programmable Respiratory Motion Software | V 4.1.20 | 2022 | Software for Motion Platform and Shaker |
Device | Type | S/N | Since | Remarks |
---|---|---|---|---|
Wyler Clinotronic PLUS | +-45° | L2407 | 2008 | Precision inclinometer for machine QA (resolution: 0.01°) |
Wyler Clinotronic PLUS | +-45° | Y1111 | 2021 | Precision inclinometer for machine QA (resolution: 0.01°) |
Wyler Clinomaster | 07/927884.1 | 2008 | Device for precise recalibration of Clinotronic PLUS | |
Wyler Clino 2000 | +-45° | K4483 | 2008 | Precision inclinometer for machine QA (resolution: 0.002°) |
Device | Type | S/N | Since | Remarks |
---|---|---|---|---|
Lufft Precision Barometer | 98849 | 1996 | Precision aneroid barometer + thermometer for reference dosimetry | |
Newport Omega DPI 740 | DPI740 | 74002599 | 2006 | Precision digital barometer (± 0.15 mbar) for reference dosimetry |
Fluke Hart Scientific 1504 | 1504 | A68539 | 2006 | Precision thermometer for reference dosimetry |
Thermistor probe | 5610-6 | 23009-1 | 2006 | Thermistor probe for use with 1504 |
Fluke "DewK" | 1620A | B08805 | 2010 | Two Channel Thermo-Hygrometer |
Thermo-Hygrometer Probe | 2626-S | B08989 | 2010 | Probe for DewK (Calibration tolerance: Temp ± 0.25°C, Humidity ±2.0 %RH) |
Thermo-Hygrometer Probe | 2626-S | B08990 | 2010 | Probe for DewK (Calibration tolerance: Temp ± 0.25°C, Humidity ±2.0 %RH) |
Thermo-Hygrometer Probe | 2626-S | B97123 | 2019 | Probe for DewK (Calibration tolerance: Temp ± 0.25°C, Humidity ±2.0 %RH) |
Device | Type | S/N | Since | Remarks |
---|---|---|---|---|
pehamed CD Mon | L991263 (via PTW) | 193 | 2010 | For acceptance testing and QA of computer monitors |
Device/Software | Type/Version | S/N | Since | Remarks |
---|---|---|---|---|
Diode Mate unit | T20010 | 00100 | 2010 | Mobile support for 12 in-vivo diodes |
Diode Mate unit | T20010 | 00101 | 2010 | Mobile support for 12 in-vivo diodes |
In-vivo diodes 6MV | various | various | 12 diodes (Sun Nuclear and PTW) | |
In-vivo diodes 15MV | various | various | 12 diodes (Sun Nuclear and PTW) | |
Isorad Electrons | various | various | 2001 | 6 diodes |
Bladder probe | T9113 | 0545 | 2005 | single detector for bladder dosimetry |
Rectum probe | T9112 | 0560 | 2009 | 5-fold detector for dosimetry in the rectum |
MultiSoft | V 1.3 | In-vivo dosimetry software |
Device/Software | Type/Version | S/N | Since | Remarks |
---|---|---|---|---|
DIAVOLT Universal | T43014 | 1352 | 2010 | Measures max and mean kVp, PPV (practical peak voltage) |
DiaControl expert | V 1.1 | 2010 | Software for evaluation of DIAVOLT measurements | |
Normi 13 | T42023 | 0550 | 2010 | Phantom for kV image QA |
Diados | T11003 | 1362 | 2004 | verified and calibrated acceptance test dosemeter |
Diados detector | TM60004 | 3356 | 2018 | verified and calibrated semiconductor for 40 - 150 kV |
CT Head & Body Phantom | T40027 | 2010 | PMMA phantoms with 16 cm and 32 cm diameter | |
CT chamber | TM30009 | 0770 | 2010 | CT chamber 3.14 ccm for DLP measurements, 100 mm |
CT Adapter | T16018 | 0479 | 2010 | Adapter for chamber HV |
Shadow Free Chamber | TM34060-2,5 | 0217 | 2010 | SFD Shadow Free Ionization Chamber 75ccm |
99.9% Cu-Filter Set | T43009.1.910 | 2010 | 13 filters from 0.01 mm to 10 mm Cu (total 21.269 mm) | |
99.99% Al-Filter Set | L981957 | 2010 | 8 filters from 0.1 mm to 10 mm Al (total 15.859 mm) | |
Stand for HVL meas. | T20008 | 2010 | holds filter and chamber | |
Catphan 600 | 2010 | Phantom for CT image QA on AquilionLB CT scanner | ||
Catphan 504 | 2012 | Phantom for CBCT image QA on TrueBeam STx | ||
PTW Conny | 11001 | 1189 | 1996 | Diagnostic dosemeter for constancy checks (30 kV, 70/100 kV) |
Dosemeter | Type | S/N | Since | Last Verified | Remarks |
---|---|---|---|---|---|
Automess Szintomat | 6134 A-H | 141578 | 2011 | 15.06.2021 | measures H*(10) |
Berthold UMo LB123 | LB123 | 122848-2138 | 1996 | 13.06.2022 | |
Berthold LB 1236 | LB1236 | 124469-1661 | 1996 | 13.06.2022 | probe used with LB123 |
FAG Eberline | FH 40 F1 | KA5021 | 1996 | 14.06.2022 | |
Graetz | GPD100 | 101391 | 1996 | 14.06.2022 | |
Graetz | GPD100 | 101392 | 1996 | 14.06.2022 | |
Graetz | GPD100 | 101394 | 1996 | 15.06.2018 | |
Graetz | GPD100 | 101395 | 1996 | 15.06.2018 | |
Graetz | GPD100 | 101396 | 1996 | 14.06.2022 | |
automess | SEQ 5 - 2 mSv | 328810 | 1996 | 15.06.2021 | |
automess | SEQ 5 - 2 mSv | 328811 | 1996 | 15.06.2021 | |
automess | SEQ 5 - 2 mSv | 328812 | 1996 | 14.06.2022 | |
automess | SEQ 5 - 2 mSv | 328813 | 1996 | 12.06.2017 | |
automess | SEQ 5 - 2 mSv | 328814 | 1996 | 14.06.2022 |
Device | Type | S/N | Since | Remarks |
---|---|---|---|---|
ALMO 3 | 100 | 2003 | Radiation Survey of treatment vault TrueBeam_BLAU | |
ALMO 3 | 099 | 2003 | Radiation Survey of treatment vault TrueBeam_GRUEN |
Device | Type | S/N | Since | Remarks |
---|---|---|---|---|
Berthold LB6411 | Id. 82842 LB6411 | 6644 | 2010 | Neutron probe |
Berthold UMo LB123 | T11003 | 1362 | 2010 | Universal monitor for Neutron probe |
Software | Version | Since | Remarks | |
---|---|---|---|---|
EPIQA | 5.4.005 | 2024 | Software for VMAT/IMRT pretreatment verification and Linac QA | |
MATLAB R2024b | 24.1 | 2024 | General purpose analysis software | |
DIAMOND | 6.2.7 | 2019 | Independent plan verification software for 3D-CRT, IMRT and VMAT | |
Track-It | 3.0 | 2020 | QA documentation and management software | |
AutoQA Plus | 1.8.6 | 2023 | Automatic evaluation of CT and CBCT image stacks | |
VERIQA | 2.1 | 2024 | Automatic independent MonteCarlo 3D dose calculation |