Farmer Chamber
Model/Serial: 30001-811 (purchased: 1996)
Model/Serial: 30001-812 (purchased: 1996)
Model/Serial: 30010-002591 (purchased: 2014)
Model/Serial: 30010-002592 (purchased: 2014)
Description: Verified ionization chamber (0.6 cm3) for absolute dosimetry. We use this non-waterproof chamber with Unidos or Unidoswebline in the home-made gantry-mounted PMMA phantom to check linac dose output during weekly linac QA.
Calibration History: Chamber 811 + Unidos 10396
Date |
NW |
U |
ppol (60Co) |
ps (60Co) |
NW' [Gy/C] |
Laboratory |
26.03.1996 |
1.00 |
0.04 |
|
|
5.132E+07 |
PTW |
20.05.1996 |
1.024 |
0.015 |
|
|
5.132E+07 |
BEV |
03.08.1998 |
1.03 |
0.02 |
|
|
5.132E+07 |
BEV |
08.03.2001 |
1.03 |
0.02 |
|
|
5.132E+07 |
BEV |
03.02.2003 |
1.03 |
0.01 |
|
|
5.132E+07 |
BEV |
05.04.2005 |
1.030 |
0.01 |
|
|
5.132E+07 |
BEV |
29.11.2007 |
1.034 |
0.014 |
|
|
5.132E+07 |
Seibersdorf |
29.04.2009 |
1.030 |
0.014 |
|
|
5.132E+07 |
Seibersdorf |
27.04.2012 |
1.032 |
0.014 |
|
|
5.132E+07 |
Seibersdorf |
15.04.2014 |
1.035 |
0.014 |
|
|
5.132E+07 |
Seibersdorf |
Calibration History: Chamber 811 + Unidoswebline
Date |
NW |
U |
ppol (60Co) |
ps (60Co) |
NW' [Gy/C] |
Laboratory |
20.04.2006 |
1.030 |
0.014 |
|
|
5.132E+07 |
Seibersdorf |
13.06.2008 |
1.030 |
0.014 |
|
|
5.132E+07 |
Seibersdorf |
06.09.2010 |
1.032 |
0.014 |
0.999 |
1.000 |
5.132E+07 |
Seibersdorf |
27.04.2012 |
1.033 |
0.014 |
1.000 |
1.000 |
5.132E+07 |
Seibersdorf |
15.04.2014 |
1.034 |
0.014 |
1.000 |
1.000 |
5.132E+07 |
Seibersdorf |
Calibration History: Chamber 812 + Unidos 10398
Date |
NW |
U |
ppol (60Co) |
ps (60Co) |
NW' [Gy/C] |
Laboratory |
26.03.1996 |
1.00 |
0.04 |
|
|
5.071E+07 |
PTW |
20.05.1996 |
1.024 |
0.015 |
|
|
5.071E+07 |
BEV |
24.11.1998 |
1.030 |
0.02 |
|
|
5.071E+07 |
BEV |
29.11.2000 |
1.030 |
0.02 |
|
|
5.071E+07 |
BEV |
06.08.2002 |
1.040 |
0.01 |
|
|
5.071E+07 |
BEV |
14.04.2003 |
1.030 |
0.01 |
|
|
5.071E+07 |
BEV |
20.04.2005 |
1.030 |
0.01 |
|
|
5.071E+07 |
BEV |
22.02.2007 |
1.025 |
0.014 |
|
|
5.071E+07 |
Seibersdorf |
20.11.2009 |
1.033 |
0.014 |
0.999 |
1.000 |
5.071E+07 |
Seibersdorf |
27.06.2011 |
1.035 |
0.014 |
0.999 |
1.000 |
5.071E+07 |
Seibersdorf |
16.04.2013 |
1.040 |
0.014 |
1.000 |
not possible |
5.071E+07 |
Seibersdorf |
Notes
- U = k uc is the expanded uncertainty with the coverage factor k = 2.
- The first calibration of the chamber was performed in Germany (PTW). Immediately after purchase, the chamber was verified and once more calibrated in Austria. The apparent "jump" in the 60-Co calibration factor Nw can be partly attributed to the different positioning rules in the two countries during the 60-Co calibration procedure.
In the figure below, the chamber is shown during calibration. In Austria (left), it is positioned with the effective point of measurement in 5 cm depth (well, not really, but a correction is made so that the effect is the same). In Germany (right), the chamber's reference point (= the chamber axis) is positioned at 5 cm depth. The result is, that the calibration factor Nw in Germany is a little bit smaller than the Austrian Nw. By how much? By the factor pdis = 1 - 0.003 r, where r is the chamber's inner radius in millimeters.
For the Farmer, r = 3.05 mm, pdis = 1 - 0.003 * 3.05 = 0.9909 or NwGER = 0.9909 * NwAUT.
Logbook
Date |
Event |
26.03.1996 |
S/N 811 and S/N 812: initial calibration by PTW. |
16.04.2013 |
S/N 812: Seibersdorf cannot determine saturation correction factor. Stabilization after voltage change takes too long. |
03.06.2014 |
Received new S/N 2601 and 2602 (cable length: 10 m). |
10.06.2014 |
Dropped S/N 2601: tip broken. |
11.06.2014 |
Both chambers sent to PTW for repair of S/N 2601 and reduction of cable length to 1.2 m. |
17.07.2014 |
Delivery of two new Farmer chambers: S/N 2591, S/N 2592 |
back to equipment