PinPoint 3D Chamber 
Model/Serial: 31016-83 (purchased: 2006)
  Model/Serial: 31016-84 (purchased: 2006, defective: 2011)
  Model/Serial: 31016-620
  (purchased: 2011)
Description: Ionization chamber 0.016 cm3. We use this detector
  for relative dosimetry in the water phantom (profiles, depth dose curves). E.g., TrueBeam depth dose curves (Eclipse beam data) for field sizes < 4x4 cm (10FFF: < 12x12 cm) were measured with this detector. 
Logbook
 
 | Date | 
 Event | 
 
 
 | 30.03.2006 | 
 S/N 83 initial calibration by PTW: ND,W = 2.462E+09 Gy/C (U = 2.1%). | 
 
 
 
 | 30.03.2006 | 
 S/N 84 initial calibration by PTW: ND,W = 2.426E+09 Gy/C (U = 2.1%). | 
 
   
  
 | 14.04.2011 | 
 S/N 83 destroyed during operation.  | 
 
 
  
 | 01.08.2011 | 
 Received S/N 620. Initial calibration by PTW: ND,W =
   2.501E+09 Gy/C (U = 1.1%). | 
 
 
 
back to equipment