0.3 cm3 Chamber

Model/Serial: 31003-939 (purchased: 1996)
Model/Serial: 31003-940 (purchased: 1996, defective: 2003)
Model/Serial: 31013-053 (purchased: 2003)

Description: Verified ionization chamber for absolute dosimetry. We use it with Unidos or Unidoswebline in the water phantom to calibrate linac dose output for photon energies.

Calibration History: Chamber 939 + Unidos 10396

Date NW U ppol (60Co) ps (60Co) NW' [Gy/C] Laboratory
26.03.1996 1.00 0.04 9.576E+07 PTW
20.05.1996 1.02 0.015 9.576E+07 BEV
03.08.1998 1.03 0.02 9.576E+07 BEV
08.03.2001 1.03 0.02 9.576E+07 BEV
03.02.2003 1.03 0.01 9.576E+07 BEV
05.04.2005 1.03 0.01 9.576E+07 BEV
29.11.2007 1.032 0.014 9.576E+07 Seibersdorf
21.04.2009 1.028 0.014 9.576E+07 Seibersdorf
27.04.2012 1.030 0.014 9.576E+07 Seibersdorf
15.04.2014 1.029 0.014 9.576E+07 Seibersdorf
25.04.2016 1.032 0.014 9.576E+07 Seibersdorf
21.06.2018 1.035 0.014 9.576E+07 Seibersdorf
14.09.2020 1.033 0.014 9.576E+07 Seibersdorf
21.06.2022 1.034 0.014 9.576E+07 Seibersdorf
19.06.2024 1.033 0.014 9.576E+07 Seibersdorf

Calibration History: Chamber 939 + Unidoswebline 11

Date NW U ppol (60Co) ps (60Co) NW' [Gy/C] Laboratory
20.04.2006 1.027 0.014 9.576E+07 Seibersdorf
13.06.2008 1.030 0.014 9.576E+07 Seibersdorf
06.09.2010 1.032 0.014 0.999 1.000 9.576E+07 Seibersdorf
27.04.2012 1.029 0.014 0.999 1.000 9.576E+07 Seibersdorf
15.04.2014 1.029 0.014 0.999 1.000 9.576E+07 Seibersdorf
25.04.2016 1.031 0.014 0.999 1.000 9.576E+07 Seibersdorf
21.06.2018 1.034 0.014 0.999 1.000 9.576E+07 Seibersdorf
14.09.2020 1.033 0.014 0.999 1.000 9.576E+07 Seibersdorf
21.06.2022 1.034 0.014 0.999 1.000 9.576E+07 Seibersdorf
19.06.2024 1.033 0.014 0.998 1.000 9.576E+07 Seibersdorf
18.06.2026 1.034 0.014     9.576E+07 Seibersdorf

Calibration History: Chamber 53 + Unidos 10398

Date NW U ppol (60Co) ps (60Co) NW' [Gy/C] Laboratory
12.03.2003 1.000 0.022 9.767E+07 PTW
14.04.2003 1.020 0.01 9.767E+07 BEV
20.04.2005 1.020 0.01 9.767E+07 BEV
22.02.2007 1.016 0.014 9.767E+07 Seibersdorf
20.11.2009 1.016 0.014 0.998 1.000 9.767E+07 Seibersdorf
27.06.2011 1.019 0.014 0.997 1.000 9.767E+07 Seibersdorf
16.04.2013 1.022 0.014 0.997 1.000 9.767E+07 Seibersdorf
30.05.2015 1.013 0.014 0.999 0.999 9.767E+07 Seibersdorf
13.06.2017 1.020 0.014 0.998 0.999 9.767E+07 Seibersdorf
19.11.2019 1.018 0.014 0.998 1.000 9.767E+07 Seibersdorf
22.06.2021 1.021 0.014 0.999 0.999 9.767E+07 Seibersdorf
19.06.2023 1.019 0.014 0.997 0.999 9.767E+07 Seibersdorf
18.06.2025 1.023 0.014 0.999 1.000 9.767E+07 Seibersdorf

Notes

60-Co calibration factors

Logbook

Date Event
26.03.1996 S/N 939 and S/N 940: initial calibration by PTW.
2003 S/N 940 defective, replaced by S/N 53.
2026 Due to increasing costs for verification and calibration, we phased out Unidos 10396 in 2026. For the remaining electrometer/chamber combinations in use, we stopped ordering the correction factors ppol (60Co) and ps (60Co). These factors are very stable (see tables).