Farmer Chamber

Model/Serial: 30010-002591 (purchased: 2014)
Model/Serial: 30010-002592 (purchased: 2014)

Description: Verified ionization chamber (0.6 cm3) for absolute dosimetry. We use this non-waterproof chamber with Unidos or Unidoswebline in the home-made gantry-mounted PMMA phantom to check linac dose output during weekly linac QA.

Calibration History: Farmer 2591 + Unidos 10396

Date NW U ppol (60Co) ps (60Co) NW' [Gy/C] Laboratory
25.04.2016 1.009 0.014 5.371E+07 Seibersdorf
21.06.2018 1.010 0.014 5.371E+07 Seibersdorf
14.09.2020 1.008 0.014 5.371E+07 Seibersdorf
21.06.2022 1.011 0.014 5.371E+07 Seibersdorf
19.06.2024 1.009 0.014 5.371E+07 Seibersdorf

Calibration History: Farmer 2591 + Unidoswebline

Date NW U ppol (60Co) ps (60Co) NW' [Gy/C] Laboratory
25.04.2016 1.010 0.014 1.000 1.000 5.371E+07 Seibersdorf
21.06.2018 1.010 0.014 1.000 1.000 5.371E+07 Seibersdorf
14.09.2020 1.009 0.014 1.000 1.000 5.371E+07 Seibersdorf
21.06.2022 1.011 0.014 1.000 1.000 5.371E+07 Seibersdorf
19.06.2024 1.008 0.014 1.000 1.000 5.371E+07 Seibersdorf
18.06.2026 1.009 0.014     5.371E+07 Seibersdorf

Calibration History: Farmer 2592 + Unidos 10398

Date NW U ppol (60Co) ps (60Co) NW' [Gy/C] Laboratory
30.05.2015 1.007 0.014 1.000 1.000 5.371E+07 Seibersdorf
13.06.2017 1.011 0.014 1.001 1.000 5.371E+07 Seibersdorf
19.11.2019 1.011 0.014 1.001 1.000 5.371E+07 Seibersdorf
22.06.2021 1.011 0.014 1.001 1.000 5.371E+07 Seibersdorf
19.06.2023 1.012 0.014 1.001 1.000 5.371E+07 Seibersdorf
18.06.2025 1.013 0.014 1.001 1.000 5.371E+07 Seibersdorf

Notes

60-Co calibration factors

Logbook

Date Event
26.03.1996 S/N 811 and S/N 812: initial calibration by PTW.
16.04.2013 S/N 812: Seibersdorf cannot determine saturation correction factor. Stabilization after voltage change takes too long.
03.06.2014 Received new S/N 2601 and 2602 (cable length: 10 m).
10.06.2014 Dropped S/N 2601: tip broken.
11.06.2014 Both chambers sent to PTW for repair of S/N 2601 and reduction of cable length to 1.2 m.
17.07.2014 Delivery of two new Farmer chambers: S/N 2591, S/N 2592
2026 Due to increasing costs for verification and calibration, we phased out Unidos 10396 in 2026. For the remaining electrometer/chamber combinations in use, we stopped ordering the correction factors ppol (60Co) and ps (60Co). These factors are very stable (see tables).