Preparation in Eclipse |
Screenshots | ||||||||||
Open a test patient in Field Setup workspace of External Beam Planning, create a new plan with one field and the energy that shall be configured (on this page it is assumed that the energy will be 6MV).
Right-click on the Field and select "Import Optimal Fluence". The corresponding fluence file can be downloaded here (right-click to save the ZIP-file to disk. The file "PDIP-configuration.optimal_fluence" is a text file and can be opened with a text editor). |
generate phantom image |
||||||||||
RT-Chart |
|||||||||||
In RT-Chart > Parameters tab, enter a Tolerance Table that will cause you no trouble during delivery. Enter couch parameters like Couch Lng = 40 cm, Lat = Vrt = 0 cm. The couch has to be out of the way during delivery. |
enter couch parameters |
||||||||||
Time Planner |
|||||||||||
In Time Planner, give the Test patient two appointments on the linac and check him in. |
create appointment check in patient |
||||||||||
4D-Console |
|||||||||||
Treat the plans on the linac and acquire the images. Note: When the field is loaded in the first plan, the imager shall be positioned vertically at -5 cm (5 cm below isocenter). For the second plan, the imager shall be at -40 cm. Therefore, the Focus imager distances are 105 cm and 140 cm, respectively. Note: Lateral and longitudinal imager positions are always zero for Portal Dosimetry! (Note for the screenshots: I changed the name of the test patient from "zz-testbox" to "zz-PDIP8114"). |
setting up 1st field ready to beam-on 15 MV image at 105 15 MV image at 140 |
||||||||||
Review Workspace |
|||||||||||
In Eclipse or Radiation Oncology, switch to Review (Dosimetry Review) and navigate to the sessions with the acquired images. Select the measured dose ( ) behind the image (expand the "+") and right-click "Eport dose to ASCII". Save it to a known location and name it "6MV_105cm.dxf". Export the second image and name it "6MV_140cm.dxf". Take care that the image named "105cm" is really the one acquired at 105 cm distance from focus! Here are some test images for 6 MV and 15 MV. Use them at your own risk. |
|||||||||||
Output Factors | |||||||||||
The output factor tables are used when the PDIP kernels are calculated. For the output factor measurements, prepare another test patient, this time in RT-Chart alone. For each field of the test plan, an image is acquired. The center pixel value of each field with reference of the center pixel value of the 10x10 field is taken as the output factor. Note that some fields with 38 cm require collimator rotation because the imager orientation is landscape (40x30 cm). It does not make a difference whether the majority of fields (which fit on the detector) have collimator rotated to 90° or are at 0°. But for some fields the collimator has to be rotated, otherwise the output factor table cannot be filled completely. It is recommended, but not required, to measure all (X,Y)-combinations of the sizes 3, 5, 10, 15, 20, 28, 38cm, except for (38x38), which is not possible. This would give 48 fields per energy. The test patient in the screenshots only has 16 fields. The following tables show our output factors for 6 MV and 15 MV, measured with 300 MU/min at 105 cm on an IDU11 detector with 50 MU per image, evaluated in Dosimetry Workspace (Point Dose in the field center). Collimator was at 0° for most fields, and rotated to 90° for Y=38 fields. |
field list in reference points tab plan scheduling sequence images collimator rotation for longest fields |
||||||||||
The tables can be downloaded here and here. Use at your own risk! | |||||||||||