Downtime Details Clinac Green (S/N 156)

Date Hours Reason for Downtime
27.02.1997 3.0 CARR, FOIL
30.05.1997 1.0 UDR1, UDR2 (bad connection on Aux backplane)
11.06.1997 2.5 GFIL
12.06.1997 2.0 GFIL (Hot Deck cover loose)
31.07.1997-
05.08.1997
32.0 Overtemp, heat exchanger blocked by dirty water from supply
15.09.1997 0.5 BMAG, XDRS
16.09.1997 2.0 GFIL "HD A/D+"
22.09.1997-
23.09.1997
6.0 Intermittent GFIL "HD A/D+"
29.10.1997 3.0 GFIL, metal span betw. gun driver chassis & optical I/F
10.12.1997 1.5 TARG
11.12.1997 4.0 TARG, changed air solenoid valve
17.12.1997 1.5 MLC: SFB-I/L, changed SF-Board
1997 59.0
14.05.1998 0.5 R-Arm does not retract
08.06.1998 6.0 HVCB, diodes blown
22.06.1998 4.0 HVCB, diodes blown
01.07.1998 4.0 HVCB, diodes blown
27.07.1998 3.0 Console computer not starting
10.08.1998 8.0 Linac temp. too low (<15°), temp. sens. defect
02.09.1998 1.0 CARR, FOIL
03.09.1998 5.5 Mirror broken
04.09.1998 1.5 CARR, FOIL, replaced switches
14.10.1998 2.0 MLC
22.12.1998 1.0 * field service report not readable *
1998 36.5
11.01.1999 1.83 CARR, FOIL
08.02.1999 2.5 HVOC, VSWR, HVCB
14.05.1999 1.5 CARR, FOIL (pin does not retract)
18.05.1999 0.3 CARR, FOIL (pin does not retract)
28.05.1999 1.0 HVCB, replaced diodes
07.06.1999 0.5 HVCB, replaced diodes
19.07.1999-
21.07.1999
23.0 Standby during beam on, burnt Stand motherboard. HVCB (diodes) SF6 leak
17.08.1999-
19.08.1999
29.5 HVOC, HVCB, troubleshooting in modulator (THY backplane etc.)
22.09.1999-
24.09.1999
20.0 HVCB, STPS, troubleshooting in modulator (THY backplane etc.)
28.09.1999 13.5 PNDT, side panels on ETR couch
04.11.1999-
15.11.1999
35.2 HVCB, troubleshooting in modulator, System Performance Audit
1999 128.83
26.04.2000 0.5 UDR1, UDR2
27.04.2000 0.5 UDR1, UDR2
03.05.2000 9.0 KSOL, replaced Klystron Solenoid
03.07.2000 2.0 R-Arm: Hands motor does not move, mushroom pin activated
27.11.2000 0.5 Sometimes ENSW in energies other than 6X
29.11.2000 4.0 Sometimes ENSW in energies other than 6X
2000 16.5
29.01.2001 1.0 CARR, FOIL
26.02.2001 5.0 BMAG
08.03.2001 4.5 CARR, FOIL, pin in/out microswitch defect
26.03.2001 1.0 GFIL, replaced Hot Deck filament PS PCB
30.03.2001 7.0 FLOW, replaced all Gantry hoses
21.05.2001-
22.05.2001
3.0 6X dose rate
03.09.2001-
05.09.2001
16.0 Collimator does not rotate, replaced motor
17.09.2001 1.25 MLC, A2
21.09.2001 6.0 MLC
01.10.2001 5.0 FLOW, replaced Stand hoses
2001 49.75
29.04.2002 1.0 Capacitor on PVI transformer
08.05.2002 0.5 MLC
08.05.2002 13.0 HVOC
31.05.2002 2.0 HVCB, Thyratron
04.06.2002-
05.06.2002
4.5 MLC, UDRS
06.06.2002 6.5 no dose rate, changed Thyratron
10.06.2002 6.5 MLC, replaced SFB
16.08.2002 4.0 Mirror
02.09.2002 0.5 MLC
2002 38.5
21.01.2003 2.5 MLC B25
23.01.2003 0.5 MLC A9
28.02.2003 0.5 MLC A9
03.03.2003 0.5 MLC A1
24.03.2003 0.5 MLC A16/A28
28.03.2003 3.5 Couch lng loose
04.04.2003 0.5 MLC A12
17.04.2003 0.5 MLC
13.06.2003 8.0 Water leak rotary joint
06.08.2003 2.0 HVOC, replaced Main Thyratron
13.11.2003 2.5 MLC B2
2003 21.5
09.01.2004 0.33 MLC B20
12.01.2004 0.33 MLC B20
28.01.2004 2.0 MLC A29&B40
18.05.2004 1.5 BMAG
14.06.2004 3.5 GFIL
19.07.2004 5.0 GAS (magnet switch)
18.08.2004 2.5 Line 2 fuse
2004 15.16
15.03.2005 0.33 GFIL (8:00-8:20)
26.04.2005 0.83 GFIL (6:30-7:20)
11.05.2005 3.5 UDRS (no output); replaced Main Thyratron and DeQing Thyratron
30.05.2005 2.0 water pump replaced
27.06.2005 1.5 FLOW due to overtemp
28.06.2005 3.5 Heat exchanger replaced (poor flow through external circuit)
20.07.2005 0.25 FLOW, PUMP: Time delay
28.07.2005 0.5 FLOW, PUMP: Time delay
19.08.2005 0.25 Line 1 Fuse
29.11.2005 4.25 CARR interlock, replaced switch
02.12.2005 1.0 ENSW
2005 17.91
04.01.2006 2.0 ENSW
09.01.2006 0.5 Noise from water pump
20.01.2006 9.0 UDRS (modulator problems)
06.02.2006 6.0 UDRS (modulator problems)
09.02.2006 2.0 ENSW
16.02.2006 4.0 UDRS
31.03.2006 4.5 Couch Lng
14.06.2006 8.0 Aux Backplane
16.06.2006 8.0 Aux Backplane
19.06.2006 8.0 Aux Backplane
21.07.2006 0.75 Collimator chain loose
21.08.2006 4.0 UDRS (modulator problems) - Thyratron replaced
16.11.2006 1.5 AFC Switch in wrong position
05.12.2006 2.0 MLC-Controller cooling problem
11.12.2006 3.5 MLC-Controller: "operating system not found" (hard disk)
21.12.2006 0.5 Water undertemp
2006 64.25
18.01.2007 4.0 HVCB, relay in Modulator
21.03.2007-
23.03.2007
26.0 ION Chamber (Transverse Symmetry)
08.06.2007 1.0 MLC (replaced hard disk)
06.08.2007 1.0 MLC
20.08.2007 1.5 GFIL
30.10.2007 1.0 Couch longitudal movement blocked
05.12.2007 4.0 Couch vertical motor
2007 38.5
28.03.2008 0.25 Collim. readout off by 2.77° (reason unknown). Recalibrated.
11.04.2008 0.25 MLC CarrA motor driver board: cable loose, 2 pins broken
15.04.2008 2.5 HWFA primary/secondary mismatch Gantry
05.05.2008 0.5 FOIL, CARR
11.06.2008 8.0 No beam
08.07.2008 2.0 GFIL
31.07.2008 0.5 Coll readout wrong
19.08.2008 1.0 HWFA (Gantry? Y1? Key?)
20.08.2008 8.5 GFIL (hot deck backplane and power supply)
21.08.2008 9.0 GFIL (TNT aircraft could not open cargo door)
22.08.2008 4.5 GFIL (TNT shipping arrived 8:40)
17.09.2008 0.5 HWFA, Line 1 fuse
19.09.2008 1.25 MLC Crg Pos
05.12.2008 4.0 MLC SF-Board
2008 42.75
14.01.2009 0.5 Console Computer "hangs"
02.04.2009 1.5 FLOW/PUMP after beaming-on for 1.5 hrs. "Pump"-fuse in Modulator.
02.05.2009 7.0 PUMP fuse in Modulator
04.05.2009 6.5 CARR/FOIL code 91 (pin did not seat)
20.05.2009 4.0 Collimator chain loose after collision Gantry-Couch
16.06.2009 0.33 MLC (Crg A)
25.06.2009 0.33 GFIL
07.08.2009 1.5 UDRS
07.08.2009 1.0 UDRS
24.08.2009 2.0 MLC
14.12.2009 4.0 Leaves on A-Side get intermittent impulses. Replaced new MLC controller by old MLC controller.
2009 28.66
19.01.2010 1.0 Some waveguide mounting parts are tumbling around freely inside the treatment head.
10.03.2010 2.0 Shielding lead block (8x5x2cm) tumbling around freely inside the treatment head (pic1, pic2, pic3).
12.03.2010 4.0 No beam. Replaced thyratron.
31.03.2010 3.0 Collimator redout wrong by 4.5° - replaced potentiometer.
21.04.2010 3.5 No power on dedicated keyboard.
21.07.2010 3.0 VAC1
28.09.2010 3.5 GFIL, Power Supply
12.10.2010 2.5 Stand Power
10.11.2010-
11.11.2010
22.0 ION Chamber replaced (follows ambient pressure). Beam adjustment with STARCHECKmaxi.
23.12.2010 0.75 GFIL
2010 45.25
07.02.2011 1.33 MLC
02.03.2011 1.25 MLC
15.04.2011 1.5 MLC
31.05.2011 2.0 MLC (replaced IR emitter)
30.06.2011 2.0 ENSW
05.10.2011 5.5 FLOW/PUMP. Replaced heat exchanger
10.10.2011 2.0 MLC (SFB)
2011 15.58
23.01.2012 1.0 Key switch
21.06.2012 1.5 Replaced IDU20 readout module
21.06.2012 0.5 FLOW, PUMP, ...
22.06.2012 0.5 MLC
06.07.2012 0.5 BMAG 9 MeV electrons
10.07.2012 0.5 Data writeback to ARIA server failed
10.07.2012 0.5 Data writeback to ARIA server failed again
20.07.2012 0.75 Couch rotation not possible over full rotation range - couch base has contact with floor
03.09.2012 0.75 GFIL
11.09.2012 0.67 Data writeback to ARIA server failed
18.09.2012 0.5 Mirror broken
20.09.2012 3.0 UDRS (DR < 150MU/min)
23.10.2012 0.67 no image acquisition possible
23.10.2012 1.5 Replaced image readout module of IDU20
04.12.2012 1.0 MLC
20.12.2012 0.5 GFIL
2012 14.34
14.01.2013 11.75 Water leak near rotary joint
2013 11.75