eMC - The Homogeneous Set
 

Standard Applicator Inserts

The first task is simple but time-consuming: calculating all combinations of energy and applicator with the default inserts that come with the applicators (not home-made inserts) using the standard SSD of 100cm. The goal is to check relative dose distributions and absolute dose (MU), with the calculation settings given above.
The planning is easy because the field sizes are exact (except for the corners, which have 8mm triangles filled out). This can be seen in the following calculation(blue) and measurement (red) of isodoses parallel to the surface in 20mm depth, for a 20MeV field and field size 4x4cm:

It is interesting to see that for 20MeV, the low dose region outside the field (still under the cerrobend cutout material) is a little underestimated by eMC: the measurement (red) is higher. The reason could be in the measurement (higher sensitivity of the electron diode for low-energy scatter) or the algorithm (IPS model, see eMC1.htm). All our home-made cutouts (see below) are molded a little thicker, so the verification of these could answer the question.

Anyway, taking into account that this is the smallest field size we allow and the highest energy, the agreement here is still very good:
The following table shows the number of particle histories that had to be processed until 1% accuracy was reached:
Particles
E06
E09
E12
E16
E20
4x4orig
6910000
12010000
22690000
41720000
63180000
6x6orig
7180000
9780000
16240000
27040000
38820000
10x6orig
11090000
14430000
25010000
40970000
58940000
10x10orig
15920000
22260000
37470000
61560000
85770000
15x15orig
37990000
48910000
78560000
122860000
272610000
20x20orig
64580000
86480000
138720000
213380000
295290000
25x25orig
100780000
138010000
209790000
316440000
447070000

(Values in bold were recalculated with Eclipse 8.1 using the Distributed Calculation Framework with a parallelization factor between 6 and 12).

The calculation times given in the next table are taken from the "Calculation Log". They are realistic in the sense that the calculations were done mid-day, under normal clinical conditions, between all other photon calculations. Since the workload on the calculation servers is not constant, there may be large variations in calculation time, for the same field.

Conditions for calculation were: 3D-image with 300 slices (slice thickness 1mm), grid 2.5mm, 1% accuracy, low 3D-gaussian smoothing.

Calc.Time [s]
E06
E09
E12
E16
E20
4x4orig
87
138
194
274
433
6x6orig
301
436
1175
1069
1827
10x6orig
942
834
3686
4901
3468
10x10orig
1266
1048
2027
5585
7140
15x15orig
2590
3589
3882
7263
1687
20x20orig
5917
6198
24585
14661
16368
25x25orig
4909
7844
20059
17513
25739
Dividing the two above tables gives the number of particle histories per second (note the increase in performance for the DCF-calculated fields in bold):
Histories/second
E06
E09
E12
E16
E20
4x4orig
79425
87029
116959
152263
145912
6x6orig
23854
22431
13821
25295
21248
10x6orig
11773
17302
6785
8360
16995
10x10orig
12575
21240
18485
11022
12013
15x15orig
14668
13628
20237
16916
161595
20x20orig
10914
13953
5642
14554
18041
25x25orig
20530
17594
10459
18069
17369
 
Home-Made Applicator Inserts

For the home-made inserts, which are usually rectangular or round, having all kinds of field sizes and which are sometimes asymmetric, care has to be taken when planning the field in Eclipse. The actual field sizes in isocenter somtimes differ a little (up to 3mm) from the nominal field sizes. This is due to the molding process. To avoid misinterpretation of the results, it is therefore a good idea first to do the measuement, determine the actual field size, and then calculate the field with the field size determined in the water phantom.

The following images show a typical comparison of a calculated (blue) and a measured (red) crossplane distribution along the depth axis, for a 4x8cm asymmetric applicator insert and 12MeV. The Gantry is at 0°:

Rotating the Gantry to 25° gives oblique incidence on the water surface (9MeV, SSD100, 6x6orig insert):
The Gamma Index is used here to evaluate the agreement. The profiles cut vertically and horizontally through the "failed" region (blue is the calculation):
The gamma criterium used was 2mm/2%, but with a suppression of doses below 5%:
 

back to Beam Verification Set